PRODUCTS...

ASTRACHECK PLUS BLOOD GLUCOSE METER

PRODUCTS / Glucometers / ASTRACHECK PLUS BLOOD GLUCOSE METER

Astracheck Plus Blood Glucose Meter

Welcome

Thank you for selecting the Astracheck Plus MM600 BGMS (Blood Glucose
Monitoring System). Follow steps for using the system in this manual.
Before using your monitoring system for the first time, please read this
manual carefully.

Intended Use

This Astracheck Plus MM600 Blood Glucose Monitoring System is for the
quantitative measurement of the concentration of glucose in whole
blood by diabetic patients or healthcare professionals as an aid in the
management of diabetes. It is intended for use outside of the body(in
vitro diagnostic use), and not intended for use on neonates.

Applicaton

Astracheck Plus MS-1D Test Strips are for in vitro diagnostic use only.
The Astracheck Plus MM600 Meter is not a substitute for your healthcare
professional. Use the Astracheck Plus MM600 Meter in conjunction with your
healthcare program.

Test Principle

Glucose in the blood sample reacts with glucose oxidase(GOD) on the
test strip and a harmless DC electrical current is produced. This current
is measured by the Astracheck Plus Meter and displayed as your blood
glucose result. The strength of these currents changes with the amount
of glucose in the blood sample. The Astracheck Plus Meter automatically
interprets this reaction.

Special Features

  • Astracheck Plus MM600 Monitoring System is designed for home usage.
  • Accurate result in 5 seconds with using only 0.5µl blood sample.
  • Large display screen.
  • Store up to 300 test results.
  • Blood glucose measurement units are pre-set in mg/dL or mmol/L.

You cannot change the unit of measure. Use of the wrong unit of
measure may cause you to misinterpret your blood glucose level, and
may lead to incorrect treatment.